The Pressure for Ventilators

Welcome to All Sensors “Put the Pressure on Us” blog. This blog brings out pressure sensor aspects in a variety of applications inspired by headlines, consumer and industry requirements, market research, government activities, and you.

The Pressure for Ventilators

In addition to the pressure to get more ventilators, pressure is an integral part of a ventilator’s operation. Breathing involves inspiratory (inhaling) pressure and expiratory (exhaling) pressure and a ventilator has to take the user’s values into account. Peak Inspiratory Pressure or PIP is the maximum pressure inside the lungs during each inhaled breath and the normal range is 25-30 cm H2O. Positive End Expiratory Pressure or PEEP is the amount of pressure left inside the lungs at the end of a breath to keep the alveoli, tiny air sacs of the lungs, open. The normal range is 3-5 cm H2O.

The pressure inside a patient’s lungs depends on the compliance of their lungs. While the suggested range of pressures during ventilation is 20-35 cm H2O with an absolute maximum of 40 cm H2O, someone with damaged lungs may need a higher pressure.

 

Airway pressure and flow waveforms during constant flow volume control ventilation show PEEP and PIP

Airway pressure and flow waveforms during constant flow volume control ventilation show PEEP and PIP.
Source: http://rc.rcjournal.com/content/59/11/1773/tab-figures-data

With pressures below 50 cm H2O (19.7 in H2O or 4,903 Pa) for dynamic measurements, a pressure sensor designed specifically for these low pressures, such as All Sensors’ DLC, DLLR, and others, provide the required accuracy.